bcs100_alarm_found_network_failed · EQUIPMENT_ALARM
BCS100 Alarm Found, NETWORK FAILED
Causes
- Loss of communication with the height sensor or BCS100 sensor
Solutions
- In CypConfig, open the FOLLOWER (BCS100) tab and click TEST CONNECT
- If the test is successful, resume operation
- If the connection test fails, check the IP address of the BCS100 height sensor in the network devices on the computer
- Set the IP address so that the last octet differs by no more than 10 — Example: BCS100 control panel shows 10.1.1.188; computer network devices should show 10.1.1.187